Vacuum Glove Box

Dual Stainless Steel Glove Box (94" x 31" x 36") with Gas Purification System and Digital Control - EQ-VGB-10-II

Dual Stainless Steel Glove Box (94" x 31" x 36") with Gas Purification System and Digital Control - EQ-VGB-10-II

기본 정보
Product Name Dual Stainless Steel Glove Box (94" x 31" x 36") with Gas Purification System and Digital Control - EQ-VGB-10-II
Sale Price Call for Price
Product Code EQ-VGB-10-II
Quantity 수량증가수량감소
상품 옵션
 
  • VGB-10-II is a dual stainless steel glove box designed for researches in material science, chemisty, semiconductor, and related topics.    The purification cylinder is made of BASF antioxidant materials from Germany and UOP water adsorbent materials from US in order to achieve a low water and oxygen concentration of lower than 1 ppm, such that experiments can be conducted in a ultraclean and highly purified gas environment.    It is integrated with vapor pressure control system, purity control system, automatic gas purity regeneration control system, autoalarm, and system data log recording.    VGB-10-II is widely used in preparing nano materials, Li-ion battery materials, catalysts, and metallorganics.
  •  SPECIFICATIONS
  • Working Voltage
  • 220 V AC,   50/60 Hz,   Single Phase
  • Max. Power Consumption
  • 3 KW
  • Glove Box Chamber
  • Case Material :  Stainless Steel 304
    Dimensions :  2400 mm (L) x 800 mm (W) x 930 mm (H)
    12 mm thick double layer laminated safety glass with corrosion protection film for operating in a safe and easy manner
  • Chamber Condition
  • Water concentration :  < 1 ppm  (20 °C,   1 atm)
    Oxygen concentration :  < 1 ppm  (20 °C,   1 atm)
  • Leak rate
  • < 0.05 Vol% / hr
    (Depending on oxygen concentration inside the glove box)
  • Air-lock Chamber
  • Antechamber Dimensions :  385 mm (Dia.) x 588 mm (L)
    Small Chamber Dimensions (for tools) :  150 mm (Dia.) x 300 mm (L)
    Max. vacuum level :  100 Pa
  • Working Gas
  • Operating gas :  Inert gas such as N2, Ar, He
    Regenerating gas :  Mixture of H2 (10 %) and Operating gas

    ○  For purification system with only water adsorbing function, Regenerating gas will be the same as Operating gas

  • Gas Purification System
  • Automated Regeneration Process
    Automated water and oxygen removal
    Purification Cylinder :  made of BASF antioxidant materials and UOP water adsorbent materials
  • Filter System
  • Organic Solvent Filter System filled with replaceable activated carbon
    Built in HEPA filter can eliminate particle size > 0.3 μm
  • Control system
  • Siemens PLC control system with 6" touch screen display

    ○  Water concentration :  0 - 1000 ppm with an accuracy of 0.1 ppm
    ○  Oxygen concentration :  0 - 1000 ppm with an accuracy of 0.1 ppm
    ○  Pressure :  -2500 to 2500 Pa with an accuracy of 1 Pa

  • Pump
  • England EDWARDS pump is included for faster vacuum and removing moisture
    Vacuum rate :  8.4 - 12 m3/hr
  • Purification Pipeline
  • 304 Stainless Steel, KF40 Port
  • Gloves
  • A pair of NORTH butyl glove is included to provide better protection against toxic materials
  • Lighting
  • Built in fluorescent lamp
  • Slidable Shelf
  • Three slidable shelves are built 0n top of inside glove chamber with dimensions of 300 mm x 250 mm
    300 mm away from roof
  • Product Dimensions
  • 3115 mm (L) x 830 mm (W) x 1830 mm (H)  (123" x 33" x 72")
  • Optional
  • Universal terminal panel is a convenient part with varies feedthrough built-in for VGB-10 series golve boxes to transfer gas/vacuum pipelines, circuit cables,etc.   ( Please send email to mti@mtikorea.co.kr for more information)
    Dimensions :  420 x 280 mm
    Adapter :  10 Φ6 ports,   4 Φ8 ports,   4 Φ12 ports      
  • Operation Instructions
  • Warranty
  • One Year limited warranty with lifetime support
    Rusting and damage due to improper storage condition or maintenance is not covered by warranty
  • Application Notes
  • The channel between the glove box chamber and the gas purification system must be unimpeded during the operation of gas purifying
    Always check the joints, gloves, sealing elements, and pipelines for air-tightness before and during the operation
    Corrosive gases are prohibited as they will damage the water sensor and oxygen sensor
    The inlet and outlet of the Air-lock Chamber must not be opened at the same time to prevent contamination by the external gas
    Coolant water outlet must remain switched on all the time
    System check frequently to replace filters and solvent adsorbents for maintanence and maximum efficiency