Features | • Heavy-duty steel frame- • PTFE Anti-Corrosion Core prevents coin cells from being
short-circuited - • An Aluminum tray is included to put at the bottom of the
crimper for avoiding oil leakage by accident
|
Output dimension | - • O.D +0.15mm after crimping (Original O.D 19.96mm, then
output is about 20.11mm)
|
Crimping Die

| • One set of crimping die for CR2032, CR2025, and CR2016- • Die is consumable, please click the picture below to order
replacement - • 100 pcs Coin cell cases (MTI's CR2032 cases) are included
in the package for testing and immediate use - • Optional die available from product option for crimping
CR2325, CR2450, AG3, AG5 - • Note: MTI does NOT carry AG5 cell cases.
|
Pressure | • Max. 8 metric Tons of hydraulic pump with automatic pressure relief- • Recommended CR20XX series coin cell sealing pressure:
850 -1000psi - • Effective piston area: 11.34 cm^2
|
Product Weight & Dimensions
 | • With handle positioned horizontally: 280mm(L) x 280mm(W) x 333mm(H)- • Without handle: 228mm(L) x 176.8mm(W) x 333mm(H)
- • Tray: 330mm(L) x 230mm(W) (See section Maintenance &
Equipment Care below for more) - • Net Weight: 26 kg (58 lbs)
- • Shipping Weight & Size: 45 kg (100 lbs) 40" x 30" x 21"
|
Compliance | CE Certified
|
Patten Number: ZL-2011-2-0462024.0 |
Warranty | One Year limited warranty with lifetime support (Rusted and damaged parts due to improper storage condition or maintenance are not covered by warranty) |
Operation Manual & Video | 
|
Maintenance & Equipment Care
 | • Electrolytes are highly corrosive materials, therefore, the die set must be cleaned with ethanol after the completion of each experiment in order to avoid any corrosion damages. Also, turn the hydraulic valve to Lock position when not in use.- • Please be aware that MTI will not be liable for any
losses/damages from improper use, improper care & maintenance and abuse of the product. - • A 13"x9" 12 gauge commercial grade aluminum tray is included
to keep the workstation clean in case of an oil leak.
|
Application Notes |  |
Maintenance & Equipment Care

| • Electrolytes are highly corrosive materials, therefore, the die set must be cleaned with ethanol after the completion of each experiment in order to avoid any corrosion damages. Also, turn the hydraulic valve to Lock position when not in use.- • Please be aware of that MTI will not be liable for any
losses/damages from improper use, improper care & maintenance and abuse of the product. - • A 13"x9" 12 gauge commercial grade aluminum tray is
included to keep the workstation clean in case of an oil leak.
|
How to Choose MTI Crimper | • MSK-110 is hydraulic manual crimper. It is not suitable for use inside the glovebox. Mainly because there is a risk of oil leakage after long time usage- • MSK-110-S is driven by inert gases, which is suitable for use
inside the glovebox with N2/Ar and with the faster operation. - • MSK-160E is driven by electric and good for using in the
glovebox with N2/Ar. However, the crimping speed is slower than MSK-110-S. It has a compact size and can be transferred via the air-lock chamber of the glovebox. - • Attention: Please use MTI Branded Coin Cell Cases with our crimper.
We don't guarantee the crimping quality if other brand cells are used.
Msk-110 MSK-110S MSK-160E
|