Vacuum Glove Box

Large Glove Box (122W x 90H x 76D, cm)  with Gas Purification System and Digital Control - EQ-VGB-6-LD

Large Glove Box (122W x 90H x 76D, cm) with Gas Purification System and Digital Control - EQ-VGB-6-LD

기본 정보
Product Name Large Glove Box (122W x 90H x 76D, cm) with Gas Purification System and Digital Control - EQ-VGB-6-LD
Sale Price Call for Price
Product Code EQ-VGB-6-LD
Quantity 수량증가수량감소
상품 옵션
 
  •  Personalization
Optional (needs special order and 8~12 weeks lead time is expected) Item Number
Glove box comes with heatable Ø360×600 antechamber  (+   ) EQ-VGB-6-LD
Glove box comes with heatable Ø150×600 small antechamber  (+   )
Glove box comes with openable main chamber window  (+   )
  • EQ-VGB-6 is a stainless steel glove box designed for researches in material science, chemisty, semiconductor, and related topics.    The purification cylinder is made of BASF R3-11 Copper Catalyst and Lindy Molecular Sieve in order to achieve a low water and oxygen concentration of lower than 1 ppm, such that experiments can be conducted in a ultra-clean and highly purified gas environment.    It is integrated with vapor pressure control system, purity control system, automatic gas purity regeneration control system, and system data log recording.    EQ-VGB-6 is widely used in preparing nano materials, Li-ion battery materials, catalysts, and metallorganics.
  •  SPECIFICATIONS
  • Working Voltage
  • 220 V AC,   50/60 Hz,   Single Phase
  • Glove Box Chamber
  • Case Material :  Stainless Steel 304 with 3 mm thickness
    Dimensions :  1220 mm (L) x 760 mm (W) x 900 mm (H)
    Removable Front Window Panel
  • Chamber Condition
  • Water concentration :  < 1 ppm  (20 °C,   1 atm)
    Oxygen concentration :  < 1 ppm (20 °C,   1 atm)
  • Leak rate
  • < 0.05 Vol% / hr
    (Depending on oxygen concentration inside the glove box)
  • Air-lock Chamber
  • Antechamber Dimensions :  360 mm (Dia.) x 600 mm (L)
    Small Chamber Dimensions (for tools) :  150 mm (Dia.) x 300 mm (L)
  • Working Gas
  • Operating gas :  Inert gas such as N2, Ar, He
    Regenerating gas :  Mixture of H2 (5 - 10%) and Operating gas

    ○  For purification system with only water adsorbing function, Regenerating gas will be the same as Operating gas

  • Gas Purification System
  • Automated Regeneration Process
    Automated water and oxygen removal
    Purification Cylinder :  made of BASF R3-11 Copper Catalyst and Lindy Molecular Sieve
  • Filter System
  • Inlet and Outlet filter can eliminate particle size > 0.3 µm
  • Pressure Control System
  • Automatic pressure control by PLC for the Glove Box Chamber

    ○  Pressure Range :  +/- 12 mbar

    Manual pressure control is allowed via a foot pedal
  • Purging System
  • Automatic Purging System allows easy operating and reduce gas consumption
  • Control system
  • Siemens PLC control system with 6" touch screen display

    ○  Water concentration :  0 - 1000 ppm with an accuracy of 0.1 ppm
    ○  Oxygen concentration :  0 - 1000 ppm with an accuracy of 0.1 ppm
    ○  Pressure :  -2500 to 2500 Pa with an accuracy of 1 Pa

  • Data
  • Automatic system data logging
  • Pump
  • England RV-8 EDWARDS pump is included for faster vacuum and removing moisture
    Vacuum rate :  8 m3/hr
  • Purification Pipeline
  • 304 Stainless Steel,   KF40 and KF25 Port
  • Gloves
  • A pair of 9 3/4" butyl glove is included for immediate use
  • Lighting
  • Built in fluorescent lamp
  • Product Dimensions  
  • 2300 mm (L) x 1500 mm (W) x 1900 mm (H)  (91" x 59" x 75")
  • Warranty
  • One Year limited warranty with lifetime support
    Rusting and damage due to improper storage condition or maintenance is not covered by warranty
  • Application Notes
  • The channel between the glove box chamber and the gas purification system must be unimpeded during the operation of gas purifying
    Always check the joints, gloves, sealing elements, and pipelines for air-tightness before and during the operation
    Corrosive gases are prohibited as they will damage the water sensor and oxygen sensor
    The inlet and outlet of the Air-lock Chamber must not be opened at the same time to prevent contamination by the external gas
    System check frequently to replace filters and solvent adsorbents for maintanence and maximum efficiency
    Note :  Customer shall order working gas (5N purity N2) and regenerating gas (N2 + 5% H2) in local to run the system
    Note :  You need at least 2 persons to remove the front window Panel due to its heavy weight and fragile structure
    Please click the image below to see EQ-VGB-6 well combined with MTI's vacuum sealer for battery R&D:   
  • Optional
    (Can be selected from option bar with extra cost)
  • Glove box comes with heatable Ø360 × 600 antechamber
     ( Please refers to the picture bottom left )
    Glove box comes with heatable Ø150 × 600 small antechamber
     ( Please refers to the picture bottom middle )
    Glove box comes with openable main chamber window
     ( Please refers to the picture bottom right )         
  • Shipping Dimension
  • 85" x 45" x 90"
  • Shipping Weight
  • 640 kg  (1400 lbs)