GF-20A Unit connects two Gas Purifier ( one for de-oxygen and another for de-humidity ) together by stainless steel pipe and two valves and ready to install to any MTI gas mixing system. It reduces gaseous impurities to sub-ppm levels at room temperature operation. It removes H2O, O2, CO, CO2, H2 and NMCH's to < 0.1 ppm from any 99.9% purity inert gas (Nitrogen, Argon, Helium, Krypton, Neon and Xenon) and H2O, O2and CO2 from Hydrogen.
Features
•Input gas purity requirement: < 1000 ppm ( 99.9%) •Output : Removal of impurities to < 0.1 ppm ( O2 ≤ 0.1ppm H2O ≤ -76oC in dew point, ( Please click here to convert humidity unit ) •No heaters or power required •Room-temperature operation •304L SS (< 10 Ra EP) vessel •Easy to regenerate •High Flow - up to 10 L/minute •Can purify inert gas upto 6 standard tanks •The filter is regenerate at 250 oC ( using 5N purity gas of 95%N2+5%H2 inside a drying oven), cusotmer can use MTI's OTF-1200X-SNT-110 compact furnace to do the regeneration (chemcial reactivation) as below: