RC-Ni is reactor made of Ni-Base Super-alloy, which has excellent creep strength and anti-oxidation.
It can reach 1100 °C with a pressure up to 4Mpa under oxygen or inert gased.
It is an ideal tool for preparing advanced material by hydro-thermal method, especially heat treat sample under high oxygen pressure.
A solenoid valve is installed on the flange allows auto-pressure relief control.
Temp.
Control and Pressure Real-time Monitoring software is also included. (new feature since 5/12/14)
SPECIFICATIONS
Structure
•Warning : For your safety, the high pressure vessel and furnace must be placed inside an explosive-proof box or case during using under high pressure.
Never face on top flange without protection to avoid accident. •Please click picture and drawing below-left to see detail structure, and click pic right to order a Explosion-Proof Box
Working Voltage
AC 208~240V, Single Phase
Pressure Vessel
•Made of Ni-Based Super-alloy, optional at 65&100ml ( click pic 1 to order spare ) •Air cold falnge via aluminium heat radiator •Liner inside vessel with variousmaterials is avlaible upon request at extra cost •Attention : Processing Tube is consumable part, you can click picture below to order replacement or spare one •Ni-Based Superalloy High Pressure Hydro-thermal Reactor
Product Dimension
Reactor Capacity & Dimensions
( Pls select the volumn in the option table )
•55 mm OD x 20 mm ID x 214 mm height, 65 mL capacity •65 mm OD x 25 mm ID x 210 mm height, 100 mL capacity
Working Temperature & Controller
•RT - 1100 °C Max. •Digital tmperature controller with 30 segment programmable •Temperature accuracy +/- 1°C
Working Pressure
•36 MPa (5292 PSI) Max at 500°C •24 MPa (3528 PSI) Max. at 800 °C •14 MPa (2116 PSI) Max. at 900 °C •7 MPa (1058 PSI) Max. at 1000 °C •4 MPa (588 PSI) Max. at 1100°C
Pressure Monitoring and Relief
•Precision high pressure sensor (measurement Range upto 250 bar) installed on pressure vessel. •Pressure monitor can display pressure in Bar. It also allows user to define the high pressure alarm set point. •When pressure reaches the alarm set point (Max.15Mpa), pressure inside the vessel will be released via the solenoid vavle installed on the flange.
For Max.pressure >15 Mpa, please shut off the manual needle valve nearby the solenoid valve.)
PC Control
•A MTS-02 control software is included for controlling and displaying temperature and pressure by PC •Strongly suggest you order a brand new labtop with pre-installed & tested software from us at extra cost. •PC control module is compatible with Labview software. •Pressure & Temp. Real Time Mornitoring (This new feature is available since 5/12/14):
Accessories
•SS316 needle valve with 1/4 pipe connector & Pressure gauge included •Two Copper Gasket 48mm O.D are included, which is consumables •Pls click picture to order spare:
Operation Instructions
Application Notes
•Attention : A two stage pressure regulator must be installed on gas cylinder to limit the pressure to below 3 PSI for safe operation.Click here to learn the installation of a gas regulator. •Do not over tightening the flange screws as this may cause permenent damages to the flange or screws.
Please refer to the instructions about proper torque for installing CF flange bolts.
We recommend you use torque wrench to tighten the screws with the flange.