Testing Equipment for UN38.3 & UL

Dual Explosion-Proof Box for Battery Safety Test ( 20"x20"x20", 125L, UN38.3.4.7 & 8 ) - MSK-TE903-UL

Dual Explosion-Proof Box for Battery Safety Test ( 20"x20"x20", 125L, UN38.3.4.7 & 8 ) - MSK-TE903-UL

기본 정보
Product Name Dual Explosion-Proof Box for Battery Safety Test ( 20"x20"x20", 125L, UN38.3.4.7 & 8 ) - MSK-TE903-UL
Sale Price Call for Price
Product Code MSKTE903
Quantity 수량증가수량감소
상품 옵션
 

The MSK-TE903-UL Dual Explosion-Proof Steel Box provides a safe enclosure for over-charging & forced discharging of all kinds of battery cell testings based on the UN38.3 standard (38.3.4.7 & 38.3.4.8). The two separate compartments with a total of 4 feedthrough ports allow for testings of multiple batteries at the same time.

Note: MSK-TE903 was updated to MSK-TE903-UL since Jan.13,2017


SPECIFICATIONS

Power

100 - 240VAC ( 50/60Hz), 100W


Structure & Material




         


  •      •  Materials:    
    • Chamber:  1.5 mm thick brushed SS304  with tensile strength > 500 Mpa
    • Exterior:    3 mm thickness cold rolled steel with tensile strength > 400 Mpa
  •      •  Structure:
            Two anti-explosion chambers, top and bottom

  •      •  Chamber Size:  500mm x 500mm x 500mm (x2)

  •      •  Testing Cables:  Two 3 meters long cable with clamp
             are included, Max. current rating is 30A

  •      •  Ports: Each chamber has two 2" diameter holes
            reserved for electric cables of battery testing
            equipment

  •      •  Safety door:Pressure relief door is built on the back
             of each chamber. In the event of a battery explosion,
            the door will open and release pressure.
     
                          
                   
     


  •      •  Note:  The max. capacity of Li-Ion batteries to be
            tested in the chamber should be limited to 4.2V@
            40Ah for single cell  or 160Wh for battery pack
            without protection circus


Venting Fan


  •      •  Venting port with high quality exhaust fan is installed
             in back chamber which shall be connected to your
             venting system.
    (ducting and exhaust system are
            not included )

  •      •  Warning:  Must keep venting fan and exhaust system
            running during battery testing to avoid pressure
            build-up in the case of battery explosion.
  •      •  Centrifugle in-line exhuast fan with aluminum flex duct
             connection is recommended.
     
             (Click pic below to read detail)

                 


Lighting


  •      •  Anti-explosion LED light (30W) is installed inside of
            each chamber.
  •      •  Angle of illumination is adjustable upto 90°

Electrode Holder & Cables

Two sets of battery electrode clamps with 10 AWG gauge cables (30A Max.) are included for easy & quick connections to batteries

Mobility

Four mobile wheels are installed at the bottom for easy transportation


Net Weight & Dimensions


  •      •  220 kg
  •      •  L700mm x W630mm x H1800mm

Shipping Weight & Dimensions

  •      •  620 lbs
  •      •  48"x40"x88"

Battery Testers (Optional)




A smart battery analyzer for use as the source of over-charging and forced discharging is available through here:




Warranty

One year limited manufacturer's warranty with lifetime support; extended warranty available upon request at additional cost.


Demo Video


   


Application Notes