Crystal Growth System

Super-alloy High Pressure Hydro-thermal Reactor 1100°C, Optional 65ml & 100ml - GSL-1100X-RC series

Super-alloy High Pressure Hydro-thermal Reactor 1100°C, Optional 65ml & 100ml - GSL-1100X-RC series

기본 정보
Product Name Super-alloy High Pressure Hydro-thermal Reactor 1100°C, Optional 65ml & 100ml - GSL-1100X-RC series
Sale Price Call for Price
Product Code GSL-1100X-RC
Quantity 수량증가수량감소
Stock 자세히
상품 옵션
UL/CSA/CE Certifications

Vessel Volume

Personalization

 
RC-Ni is reactor made of Ni-Base Superalloy, which has excellent creep strength and anti-oxidation. It can reach 1100 °C with a pressure up to 4Mpa under oxygen or inert gas. It is an ideal tool for preparing advanced material by the hydrothermal method, especially heat treat sample under high oxygen pressure. A solenoid valve is installed on the flange allows auto-pressure relief control. Temp. Control and Pressure Real-time Monitoring software are also included. (new feature since 5/12/14)
 
SPECIFICATIONS




Structure

  •      •  Warning:   For your safety,  the high-pressure vessel and furnace must be placed inside an explosive-
            proof box or case during using under high pressure.  Never face on top flange without protection to avoid
            the accident.
      
  •      •  Please click picture and drawing below left  to see detail structure, and click pic right to order an
            Explosion-Proof Box

                             


Working Voltage
  •      •  AC 208~240V, Single  Phase, 2 KW,

Pressure Vessel



  •      • 
    Made of Ni-Based Superalloy, optional at 65&100ml ( click pic 1 to order spare )

  •      •  Air cold flange via aluminum heat radiator
  •      •  Liner inside vessel with various materials is available upon request at extra cost
  •      •  Attention: Processing Tube is consumable part, you can click picture below to order replacement or spare
            one  It is user resposibility to choose suitable liner to grow different material
  •      •  A small quartz or Alumina boat should be used to load the sample inside the vessel. Please click the link
            below to order separately.
    Never place the sample directly onto the vessel without a crucible. Vessel
           service-life could be reduced significantly if you do so.
  •      •  You may use gold or nickel foil to wrap the crucible boat to minimize the contamination to the pressure vessel.

  •                     


Product Dimension



Reactor Capacity & Dimensions
( Pls select the volumn in the option bar )

  •      •  55 mm OD x 20 mm ID x 214 mm height, 65 mL capacity
  •      •  65mm OD x 25mm ID x 210 mm height, 100 mL capacity

Working Temperature & Control




  •      •  RT -  1100 °C Max.
  •      •  Digital temperature controller with 30 segment programmable
  •      •  Temperature accuracy +/- 1°C
  •      •  Recommend heating rate <=10°C/min

Thermocouple

 


Working Pressure

65ml Version Pressure Range

Temperature Range

100ml Version Pressure Range

Temperature Range

≦21 Mpa

 ≦800°C

≦20 Mpa

 ≦800°C

≦12.5 Mpa

 ≦900°C

≦12 Mpa

 ≦900°C

≦6.3Mpa

≦1000°C

≦6 Mpa

 ≦1000°C

≦4.2Mpa

≦1100°C

≦4 Mpa

≦1100°C


Pressure SensorPrecision high-pressure sensor (measurement Range up to 250 bars) installed on the pressure vessel.

Pressure Relief Solenoid Valve

  •      • 
    A default 15Mpa max. pressure relief solenoid valve is installed. When the pressure reaches the alarm set
            point (Max.15Mpa), the pressure inside the vessel will be released via the solenoid valve installed on the
            flange. For Max.pressure >15 Mpa, please shut off the manual needle valve nearby the solenoid valve.)



PC Control (optional)

  •      •  A MTS-02 control software is available for controlling and displaying temperature and pressure by PC
  •      •  PC control module is compatible with Labview software.
  •      •  Pressure & Temp. Real Time Monitoring (This new feature is available since 5/12/14):  

                        


Accessories


  •      •  SS316 needle valve with 1/4 pipe connector & Pressure gauge included
  •      •  Two Copper Gasket 48mm O.D are included, which is consumables
  •      •  Pls click picture to order spare
  •      •  The optional torque wrench is available here  to avoid overtightening the flange bolts.Please email us to
            learn the suitable tightening torque for the Hydrothermal Reactor.

  •               


Operation Instructions

   
         


WarrantyOne year limited warranty. Consumable parts such as copper gaskets, pressure vessels (if over-pressurized, over-heated or chemical interactions), and heating elements (if over heated) are not covered by the warranty.  

Application Notes

  •      •  Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the
            pressure to below 3 PSI for safe operation.
     
    Click here to learn the installation of a gas regulator.
  •      •  Do not over tightening the flange screws as this may cause permanent damages to the flange or screws.
            Please refer to
     the instructions about proper torque for installing CF flange bolts.
    We recommend you
           use a torque wrench to tighten the screws with the flange.

Shipping Dimensions
  • 40" x 32" x30"
Shipping Weight
  • 150 lbs

炉管内最大气压

    加热温度

 

 

20 Mpa

800°C

 

 

≦12Mpa

900°C

 

 

≦6Mpa

1000°C

 

 

≦4Mpa

1100°C