Testing Equipment for UN38.3 & UL

Portable Digital Temperature Calibrator with PC Software - NIST certificated - EQ-TC-calibrator-LD

Portable Digital Temperature Calibrator with PC Software - NIST certificated - EQ-TC-calibrator-LD

기본 정보
Product Name Portable Digital Temperature Calibrator with PC Software - NIST certificated - EQ-TC-calibrator-LD
Sale Price Call for Price
Product Code EQ-TC-calibrator-LD
Quantity 수량증가수량감소
상품 옵션
 
 

The CL3515R is a portable precision calibrator/thermometer with a 4-digit LCD. It is designed to use external K/J/T/E/R/S/N/L/U/B/C type thermocouples as temperature sensors. MTI offers a complete selection of thermocouples, or we can custom-design parts to meet your specifications. The thermometer features a dual thermocouple input and an adjustable thermocouple offset.


SPECIFICATIONS




Features:

  • 11 Types of Thermocouples: K, J, T, E, R, S, N, L, U, B, and C (in oC, oF, or K), and mV Signal Output
  • USB and RS232 Interface with Windows Software (compatible with winXP,winVista and win7)
               
  • NIST-Traceable Calibration Certificate with 13 Data Points (Included)
  • 2 Channels for Thermocouple Temperature Measurement (T1 and T2) with Offset Adjustment
  • Ergonomic Rotary Knob Allows for One-Handed Operation




Thermocouple (Resolution) Measurement Range*
* According to temperature standard ITS-90.
  • K: (0.1o) -200 to 1372 oC(-328 to 2501oF) 
  • J: (0.1o) -210 to 1200 oC(-346 to 2192 oF ) 
  • T: (0.1o) -250 to 400 oC(-418 to 752 oF ) 
  • E: (0.1o) -250 to 1000 oC(-418 to 1832 oF ) 
  • R: (1o) 0 to 1767 oC(32 to 3212 oF ) 
  • S: (1o) 0 to 1767 oC (32 to 3212 oF ) 
  • N: (0.1o) -200 to 1300 oC (-328 to 2372 oF ) 
  • L: (0.1o) -200 to 900 oC (-328 to 1652 oF ) 
  • U: (0.1o) -200 to 600 oC (-328 to 1112 oF  ) 
  • B: (1o) 600 to 1820 oC (1112 to 3308 oF ) 
  • C: (1o) 0 to 2316 oC (32 to 4200 oF )



Accuracy:
  • Thermocouple Types K/J/T/E/L/U: 
    • +/-(0.05% rdg + 0.5oC) -50 to 1372oC[+/-((0.05% rdg + 1.0 oF ) -58 to 2501oF] 
    • +/-((0.05% rdg + 1.0oC) -50 to -250oC[+/-((0.05% rdg + 2.0 oF ) -58 to -346 oF ] 
  • Thermocouple Type N: 
    • +/-((0.05% rdg + 1.0oC) -50 to 0 oC [+/-((0.05% rdg + 2.0 oF ) -58 to 32 oF ] 
    • +/-((0.05% rdg + 0.5oC) 0 to 1300oC[+/-((0.05% rdg + 1.0 oF ) 32 to 2372 oF ] 
  • Thermocouple Types R/S/B/C: 
    • +/-((0.05% rdg + 2oC) 0 to 1767oC[+/-((0.05% rdg + 4oF ) 32 to 3212 oF ]



Thermocouple Simulation Range:

  • Resolution: 0.1o  (1o for R/S/B/C) 
  • Accuracy*: +/-(0.3oC + 10 uV) 
  • * Accuracy is specified for operating temperatures from 18 to 28oC (64 to 82 oF ) for 1 year, not including thermocouple error. 
  • mV Range: -25 to 75 mV 
  • Resolution: 10 uV 
  • Accuracy: +/-(0.025% + 1 digit
Temperature Coefficient:0.1 times the applicable accuracy specification per oC  from 0 to 18 oC  and 28 to 50oC (32 to 64oFand 82 to 122oF)
Display4-digit LCD with max reading of 19999
Input ConnectorMiniature thermocouple connector (SMP type) Power: 9V battery (included)
Battery Life17.5 hours nominal with low-battery indicator
Dimensions192 H x 91 W x 52.5 mm D(7.5 x 3.5 x 2.1")
Weight320 g (0.7 lb)
Environmental Ambient Operating Ranges0 to 50oC (32 to 122oF), <80% RH
Storage Ranges-20 to 60oC (-4 to 140oF), <70% RH
ManualManual ,    Software
WarrantyOne year limited warranty with lifetime support

Packing List
1Digital Temperature Calibrator1pc
2IR2 to USB connection cable1pc
3IR2 to RS232 connection cable1pc
4Thermocouple connector1pc
5Thermocouple extension cable2pcs
69V Battery1pc
7Software CD1pc
8User Manual1pc