  • RC-Ni-V is a compact reactor made of Ni-Base Super-alloy, which has excellent creep strength and anti-oxidation.    It can reach 1100 °C with a pressure up to 4Mpa under oxygen or inert gased.    It is ideal tool for preparing advanced material by hydro-thermal method, especially heat treat sample under high oxygen pressure.
  •  SPECIFICATIONS
  • Structure
  • Made of Ni-Based Super-alloy ( Waspaloy or equivalent )
  • Reactor Capacity & Dimensions (Please select the volumn in the option table)
  • 55 mm OD x 20 mm ID x 210 mm height, 65 mL capacity
    65 mm OD x 25 mm ID x 210 mm height, 100 mL capacity
  • working Temperature
  • RT - 1100 °C Max.
  • Working Pressure
  • 36 MPa (5292 PSI)   Max at 500°C
    20 MPa (2900 PSI)   Max. at 800 °C
    12 MPa (1740 PSI)   Max. at 900 °C
    6 MPa (870 PSI)   Max. at 1000 °C
    4 MPa (588 PSI)   Max. at 1100°C
  • Accessories
  • SS306 needle valve with 1/4 pipe connector included
  • Optional Upgrades
    ( click picture to review details )
  • 304SS 1/4" O.D tube fitting  - for connecting reactor to gas tankvia 1/4" SS pipe You may click the picture to order dighi-pressure gauge You may picture above to order a pressure relieft valve for safety operation
    Optional: You may insert reactor into vertical furnace with Temperature controller to 1100°C. ( click picture to order whole system )
  • Warning
  • For your safety, the high pressure vessel and furnace must be placed inside a explorsive-proof box during using under high pressure box or under fumehood.    never face on top flange to avoid accident.
    Attention : Due to the high pressure gas output from the gas cylinder, it is necessary to install the gas pressure regulator with it to reduce the output pressure, avoid overpressurizing the furance chamber and ensure a safe operation.    Click here to learn the installation of a gas regulator. 
    We recommend you use torque spanner to tighten the screws with the flange.    Proper torque should be used is 200 N⋅m.    Do not over tightening the flange screws as this may cause permenent damages to the flange or screws.