Structure
| • Thermal efficient ceramic (high purity alumina fiber) insulation surrounding the chamber provides maximum energy efficiency. - • Heating elements on all three sides of the interior chamber for fast heating and enhanced
temperature uniformity. - • Heavy duty double layer structure with the cooling fan to keep furnace exteriors safe to touch.
- • Built-in vent port removes undesirable contaminants and moisture; extends the life span of the
heating elements and furnace. - • 1/4'' Barbed Gas Inlet/Outlet installed on the back of the furnace
- • An interlock is installed to cut off power as soon as the door is open
|
Chamber Dimensions
| • 16 " x 16'' x 16'' (D x H x W) (40x40x40 cm)- • Capacity:64 liters/ 4100 cubic-inches/ 2.37 cubic ft
- • Observation window of quartz ( 1/2" - 2" diameter ) is available upon request
-
|
Power | • 7.5KW- • Power cord supplied without plug
|
Current Rating | 34A |
Voltage
| AC 208 - 240 Single phase, 50/60Hz Note: Power cable included, but without the plug. Please install a plug by your own or click link below picture to choose a plug according to your country or your lab requirement
|
Operating Temp Range
| • 100-1100°C-Max. continuously- • 1200°C-Max. under an hour
- • Note: for using =>1000°C, furnace block must be placed front chamber, see picture below.
|
Stability at 1000oC | +/- 1oC |
Uniformity at 1000oC | +/- 5oC |
Max. Heating Rate | 20°C/ min |
Temperature Controller
| • FA-YD518P-AG temperature controller is included.- • Proportional–integral–derivative control (PID control) and auto-tune function
- • 30 segments programmed with ramping, cooling and dwelling steps
- • Built-in over-temperature alarm and thermocouple failure alarm
- • +/- 1 ºC temperature control accuracy
- • Default DB9 PC communication port
- • MET Certified
|
Temperature controller (optional)
| • MTS02-Y Temperature control software kit (for YD518P series controllers) + 15" laptop package is available at the options bar. Click for the detail information about the MTS02-Y software kit.
- • You may upgrade the temperature controller to the Eurotherm Temperature Controller which is fully
compatible with LabVIEW. This package includes a Eurotherm programmer, a communication cable (RS485 - USB) and a software CD. Click the pic below to view the full spec of the Eurotherm temperature controller.
|
Laptop, software & WiFi Control(Optional) | • Brand new laptop with Microsoft Window 10 and Microsoft Office 2013 (30 days free trial) for immediate use.- • Labview Based Temperature Control System (EQ-MTS01) enables user to edit temperature profile,
manage heat-treatment recipe, record and plot data for MTI furnaces. - • A wireless remote control provides up to 300 meters operating range.
- • Above features are available upon request at an extra cost (up to $1,000). Please contact us for
more information. - • Note: The software is only compatible with MTI’s Yudian Temperature Controller
|
Furnace Weight | 275 LB |
Product Dimension | 660mm L x 645mm W x 945mm H |
Shipping Dimensions
| 48" x 40'' x 54" (DxHxW)
|
Shipping Weight | 425 Lbs (shipping by pallet) |
Warranty | • One-year limited warranty with lifetime support. - • ATTENTION: Any damages caused by the use of corrosive and acidic gasses are not under
the coverage of MTI One Year Limited Warranty.
|
Compliance | • CE Certified- • NRTL(UL61010) or CSA certification is available upon request at extra cost.
|
Application Notes | • For longer furnace service life, components, we recommend the heating rate not to exceed 10 °C /min.- • The cooling rate shall also not exceed 10 °C /min.
- • This furnace lacks vacuum sealing, toxic or explosive gasses are not recommended for use with this
furnace without necessary safety controls and supervision. - • Small cracks may appear on the surface of the refractory ceramics over extended use.
This is a normal occurrence and the cracks may be repaired with alumina coating.
- click here to choose a crucible
|
Operation Instructions |
|