GPH-DHDO connects two gas purifiers ( one for de-oxygen and another for de-humidity ) together by 1/4'' stainless steel pipe and two valves and ready to install to any MTI gas mixing system. It reduces gaseous impurities to sub-ppm levels at room temperature operation. It removes H2O, O2, CO, CO2, H2 and NMCH's to <0.1 ppm from any 99.9% purity inert gas (Nitrogen, Argon, Helium, Krypton, Neon and Xenon) and H2O, O2and CO2 from Hydrogen. What is more, it has buit-in heater which can be heated to 300C to easily regenerate the de-oxygen and de-hydration chemical in the unit for long term use.
•304L SS (<10 Ra EP) vessel •Easy to regenerate •Digital temperature controller and Gas flow-meter control are built in •Purification capacity upto 80L inert gas after regeneration •The unit should be regenerated at 270 oC ( using 5N purity gas of 95%N2+5%H2)
Specifications
Power Input
•208-240V single phase 50/60Hz (110VAC version is avaialble upon request ) •Power Consumption: 500W (when heating)
Maximum Pressure Input
0.5Mpa
Operating Temperature
Room Temperature- 300oC
Leak Rate
< 2 x 10-9 atm cc/sec He
Temperature Controller
MTE cerified Digital temperature controller with +/- 1oC accuracy