PPM Grade Dual Column Inert Gas Purifier with Regeneration Heater ( Removing O2 & H2O to PPM ) - GPH-DHDO

PPM Grade Dual Column Inert Gas Purifier with Regeneration Heater ( Removing O2 & H2O to PPM ) - GPH-DHDO

기본 정보
Product Name PPM Grade Dual Column Inert Gas Purifier with Regeneration Heater ( Removing O2 & H2O to PPM ) - GPH-DHDO
Sale Price Call for Price
Product Code GPH-DHDO
Quantity 수량증가수량감소
상품 옵션
 
  • GPH-DHDO connects two gas purifiers ( one for de-oxygen and another for de-humidity ) together by 1/4'' stainless steel pipe and two valves and ready to install to any MTI gas mixing system.  It reduces gaseous impurities to sub-ppm levels at room temperature operation.  It removes H2O, O2, CO, CO2, H2 and NMCH's to <0.1 ppm from any 99.9% purity inert gas (Nitrogen, Argon, Helium, Krypton, Neon and Xenon) and H2O, O2and CO2 from Hydrogen.  What is more, it has buit-in heater which can be heated to 300C to easily regenerate the de-oxygen and de-hydration chemical in the unit for long term use.
  •  Features
  • Input gas purity requirement
  • < 1000 ppm ( 99.9%)
  • 304L SS (<10 Ra EP) vessel
    Easy to regenerate
    Digital temperature controller and Gas flow-meter control are built in
    Purification capacity upto 80L inert gas after regeneration
    The unit should be regenerated at 270 oC ( using 5N purity gas of 95%N2+5%H2)
  •  Specifications
  • Power Input
  • 208-240V single phase 50/60Hz
     (110VAC version is avaialble upon request )
    Power Consumption: 500W (when heating)
  • Maximum Pressure Input
  • 0.5Mpa
  • Operating Temperature
  • Room Temperature- 300oC
  • Leak Rate
  • < 2 x 10-9 atm cc/sec He
  • Temperature Controller
  • MTE cerified Digital temperature controller with +/- 1oC accuracy
  • Max. Flow rate
  • 400mL/minute
  • Gas Fittings
  • 1/4 Swagelok
  • Size
  • 445 x 400 x 190mm