Electric Hydraulic Pump
| - Power: 208-240VAC, 50/60Hz
- 180 W AC motor.
- Programmable and automatic pressure control up to 70 MPa
- Accuracy: +/- 0.5MPa in high pumping speed, +/-0.1 MPa in low pumping speed
- Built-in digital 30-segment programmable pressure controller to set up multi-step pressurization and dwell time
- The PC software is included for setting up the pressure program and data logging.
|
Max. Working Pressure | Gauge Displayed Pressure Value (MPa)
| 5 | 10 | 20 | 25 | 30 | 41 | 55 | Actual Force Applied on the Hydraulic Cylinder (Ton) | 5.5 | 8 | 15 | 18.5 | 22 | 30 | 40 |
- The actual force applied to the Hydraulic Cylinder (Ton) is equal to the force applied to your working piece
- For calculating the actual pressure applied to the working piece, please follow the equations below:
- P = T/(πr^2) x g x 10
- P: Pressure (MPa) applied to your working piece
- T: Force (Ton) applied to the Hydraulic Cylinder
- g: Standard gravity. g ≈ 9.8 m/s2
- π: ~3.14
- r: Radius (cm) of the working piece
- *1 Metric Ton = 1000 kg; 1 MPa ≈ 10 kgf/cm2
|
Laminating Hot Press
| - Power: 208 - 240 VAC, 50/60Hz, single phase
- 4.5 kW max.
- 20A circuit breaker is required for 220V, 5' length power cable included, but no plug
- Platen Area: 300 mm x 300 mm (12"x12")
- Material: Harden Cr12MoV alloy
- Surface: Flattening with CNC grinding
- The parallelism between two plates: < 0.1 mm within 10"x10" area at 200°C
- Plate surface roughness: < 2 microns
|
Max. Travel Distance
| - 80 mm
- Do not raise the bottom cylinder (piston) more than its maximum traveling distance (80 mm)
|
Pressing Plate Max. Opening Height | |
Max. Working Temperature | - 200°C continuous with water cooling
|
Heating Rate & Cooling Rate | |
Water Cooling Jacket
| - Two water cooling jackets are built on the back of the top and bottom heating plate separately with 1/4" water pipe connectors. Such a configuration allows you to cool the sample as soon as the hot pressing process is completed.
- One recirculating water chiller is required but not included in the standard package.
Both the KJ5000 water chiller and the KJ5300 water chiller would be suitable for the operation. -
|
Temperature Control | - Two digital controllers are installed in one box with 30 segments programmable and auto-tune function
- Temperature accuracy ±2°C
- The PC software is included for setting up the temperature profile and data logging.
- Note: The PC software for setting up the temperature and the pressure are two different software, but both controllers can be connected to a single PC and both software can be run at the same time.
|
Compliance
| - CE Certified for the whole system
- NRTL (UL 61010) or CSA Certification is available upon request, additional charges will apply
|
Dimensions | Click the pic for more detail. |
Operation Video | Install Instruction 100 um plane parallelism testing
|
Warranty | - The pressing machine was quality inspected in the USA, lifetime support is provided
- One-year limited standard warranty (Extended warranty is available for up to 5 years at an extra cost)
|
Application Note | - The Hot Press can be placed inside of a glovebox and a temperature controller and the pump will be put outside of the glovebox (Pic. 1)
- The hot press can be used for measuring the thermo-electric generator module ( top hot and bottom cool) Pic. 2
- Pic. 1 Pic. 2
|