Split Single Zone Tube Furnaces...

UL Standard 1200C Split Tube Furnace with Vacuum Flanges & Optional Quartz Tube 60, 70, 80, or 100 mm O.D. - OTF-1200X series

UL Standard 1200C Split Tube Furnace with Vacuum Flanges & Optional Quartz Tube 60, 70, 80, or 100 mm O.D. - OTF-1200X series

기본 정보
Product Name UL Standard 1200C Split Tube Furnace with Vacuum Flanges & Optional Quartz Tube 60, 70, 80, or 100 mm O.D. - OTF-1200X series
Sale Price Call for Price
Product Code OTF-1200X series
Quantity 수량증가수량감소
Stock 자세히
상품 옵션
UL/CSA Certification

furnace size

Personalization(Upgrade to Eurotherm Temperature controller)

 
OTF-1200X-UL furnace is configurated to meet UL/CSA certificate standard. The furnace can adopt a 60, 70, 80 or 100mm O.D quartz tube and a pair of stainless steel vacuum sealing flanges with needle valves & gauge allow heating samples under vacuum or gas flow conditions. The temperature of this tube furnace is controlled by a high precision digital controller which provides 30 programmable segments with +/- 1°C accuracy. The max. working temperature is 1200°C.

SPECIFICATIONS:


Furnace Structure



  •      •  Double layer steel casing with air cooling keeps furnace surface temperature lower than 60°C.  
  •      •  High purity fibrous alumina insulation for maximum energy saving.
  •      •  The special refractory coating on liner and heating element to increase furnace life.
  •      •  Split cover with Interlock protection, cut the power if door open during furnace heating process.


Voltage and Power

Input voltage
208 - 240VAC Single Phase, 50/60Hz
3KW, Circuit breaker 20A


Heating Temperature

  •      •  Fe-Cr-Al Alloy doped by Mo
  •      •  Max. heating temperature: 1200°C
            (
    N2 or other inert gas must be filled inside the tube to resist deformation)
  •      •  Continuously heating temperature: 1100°C
  •      •  Heating rate: 20°C /min Ma
  •      •  Thermal Couples: K type 220mm (8.7") Length with Ceramic Sheath, please click the link to
              order spare.


Heating Zone

  •      •  Total heating zone length: 440mm (~ 18")
  •      •  Constant temperature zone: 150mm (+/1°C)
  •      •  Please click the image below to view the temperature distribution at 980°C.
                        Temperature Distribution


Temperature Controller



  •      •  FA-YD518P-AG
     temperature controller is included.
  •      •  Proportional–integral–derivative control (PID control) and auto-tune function
  •      •  30 segments programmed with ramping, cooling and dwelling steps
  •      •  Built-in over-temperature alarm and thermocouple failure alarm
  •      •  +/- 1 ºC temperature control accuracy
  •      •  Default DB9 PC communication port
  •      •  MET Certified


Temperature controller (
optional)


  •      •  MTS02-Y Temperature control software kit (for YD518P series controllers) + 15" laptop
            package is
     available at the options bar. 
           Click for the detail information about the MTS02-Y software kit.
  •      •  You may upgrade the temperature controller to the Eurotherm 3204 Temperature
            Controller
     which is fully compatible with LabVIEW. This package includes a Eurotherm 3204
             programmer, a communication cable (RS485 - USB) and a software CD.
             Click the pic below to view the full spec of Eurotherm 3504 temperature controller.
             
     
Fused Quartz Tube SizeOTF-1200X-60
60 O.Dx55 I.Dx1000 mm L
OTF-1200X-70
70 O.Dx65 I.Dx1000 mm L
OTF-1200X-80
80 O.Dx74 I.Dx1000mm L
OTF-1200X-100
101 O.Dx92 I.Dx 1000mm L

Refractory Blocks

  •      •  Two fibrous ceramic tube blocks are included for blocking heat radiation from inside tube.
            (Ceramic blocks must be fully inserted into the furnace chamber before heating)
  •      •  Quartz blocks (optional) are ideal for cleaner processing environment and higher vacuum
             efficiency.
  •      •  You may order flange support to avoid tube rotating ( click Pic below right )

            


Vacuum Flange




  •      •  One pair of SS 304 vacuum flange is included
  •      •  Left flange:
    •          º  one thermal feedthrough (for 1/4O.D sheathed thermocouple)
    •          º  one needle valve
    •          º  one mechanical vacuum gauge -0.1Mpa to 0.15Mpa
    •          º  one barb fitting ( click PiC, below left-2 for detail)
  •      •  Right flange:
    •          º  one KF-25 vacuum port  and one needle valve.
                                     
Note: The standard flange for 70mm O.D tube has no KF-25 port or thermocouple feedthrough. Click there to see detail pic.  

Optional
  •      •  The flange with quartz window is available for in-situ Raman or Mossbauer analysis
                


Vacuum Pressure

  •      •  10E-2 torr by quality mechanical vacuum pump
  •      •  10E-5 torr by the turbo molecular pump.
  •      •  The vacuum pump is optional. Please click here to choose a vacuum pump.
                

Product Dimensions
Furnace: 590mm(L) ×380mm(W) × 520mm(H)
Net Weight
40kg
Shipping Dimensions45"(L) x 45"(W) x 42"(H)
Shipping Weight200lbs
Warranty
One year limited warranty with lifetime support (Consumable parts such as processing tubes, O-rings are not covered by the warranty, please order replacements at related products below.)

Laptop, software & WiFi Control(Optional)

  •      •  Brand new laptop with Microsoft Window 10 and Microsoft Office 2013 (30 days free trial) for
            immediate use.
  •      •  Labview Based Temperature Control System (EQ-MTS01) enables user to edit temperature
            profile, manage heat-treatment recipe, record and plot data for MTI furnaces.
  •      •  A wireless remote control provides up to 300 meters operating range.
  •      •  Above features are available upon request at an extra cost (up to $1,000). Please contact us
            for more information.
  •      •  Note:  The software is only compatible with MTI’s Yudian Temperature Controller

  •            


Compliance

     •  CE Certificate
     •  UL ready
     •  CSA ready
     •  Click here to understand the certificate and condition.

                     


Application Notes



Click here to learn the installation of a gas regulator

  •      •  Tube furnaces with quartz tubes are designed for using under vacuum and low pressure < 0.2
            bars / 3 psi / 0.02 MPa.
    Vacuum pressures may only be safely used up to 1000°CThe flow rate
            for gasses should be limited to<200 SCCM (or 200ml/min) for reducing thermal shocks to the
            tube
  •      •  Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit
            the pressure to below 3 PSI for safe operation.
     
            Click here to learn the installation of a gas regulator.
  •      •  Heating module is consumable.  Please click picture below-right to order a replacement.
  •      •  The furnace with Sliding flange is available
                


Operation Instructions & Video


Click here to learn How to set up quartz/ ceramic tube and vacuum flange for MTI Tube furnace. ( also, click picture left to see demo video )


              


STANDARD PACKAGE:

Description
Qty
Pic
OTF-1200X-60
OTF-1200X-70OTF-1200X-80OTF-1200X-100
1
Alumina foam block for tube furnace1 pair

EQ-F-T-block-52
EQ-F-T-block-62
EQ-F-T-block-70EQ-F-T-block-92
2
Vacuum flanges
1 pair

EQ-FL-60
EQ-FL-70
EQ-FL-80EQ-FL-101KF25-FT
3
Fused quartz tubing
1 unit

EQ-QZTube-60G-1000
EQ-QZTube-70G-1000
EQ-QZTube-80D40EQ-QZTube-100