Controlled Gas Delivery

Gas Independent Automatic Pressure Control Set with Dual Solenoid Valves, 4E-5 - 1E3 Torr - EQ-PCG-2V-UL-LD

Gas Independent Automatic Pressure Control Set with Dual Solenoid Valves, 4E-5 - 1E3 Torr - EQ-PCG-2V-UL-LD

기본 정보
Product Name Gas Independent Automatic Pressure Control Set with Dual Solenoid Valves, 4E-5 - 1E3 Torr - EQ-PCG-2V-UL-LD
Sale Price Call for Price
Product Code KJT-PGC
Quantity 수량증가수량감소
상품 옵션
 

EQ-PCG-2V-LD automatic pressure control system is designed for applications that need controlling constant pressure in a closed system with an in-line flow such as Vacuum Tube Furnaces. The Anti-corrosive Digital Vacuum Gauge provides an RS-232 analog output of the pressure signal to pressure control unit which directly controls those two solenoid valves (a feed valve and an exhaust valve).  

  •      •  Only suitable for Mechanical Pumpimproperly usage the EQ-PCG-2V-LD with turbo-pump may permanently damage pump-vane.
  •      •  If you need turbo pump to build a Precision Vacuum Pressure Regulating System, please consider our Motorized Valve and Gauge.

SPECIFICATIONS
 

Complete Package includes



Pressure Range

  •      •  Display range: 3.8 x 10-5 to 1125 Torr.
  •      •  Controllable range: 1mTorr-760Torr.

Working voltageAC 110~220V





Typical Operation Procedures for Controlling the Pressure of Tube Furnace 

  •      •  Install feed valve (inlet) on one end of tube flange, then connect it with gas supply cylinder.
  •      •  Install exhaust valve (outlet) on the other end of tube flange, then connect it to vacuum pump or gas recovery
             system
  •      •  Install digital pressure gauge & controller on the tube flange
  •      •  Well connect the Pressure Gauge and two Solenoid Valves to Pressure Control Unit,
  •      •  Power on the system by 110/220 VAC power source.
  •      •  Set pressure low limit and high limit on digital pressure gauge
  •      •  The pressure inside the tube chamber will be kept within the desired range

           






Application Notes

Gas flow-in/out rate can largely affect system pressure control accuracy, the minimum range of pressure control recorded by using the system shown in below is:


         
  •      •  +/- 100 mtorr (<1 torr),
  •      •  +/- 20 torr (1~760 torr),
Please click the picture above to see details of system setup.



Operation Manual





Compliance

  •      •  Vacuum gauge is ETL certified
  •      •  All electric components in control box ( >24V) are UL certified
  •      •  CE certified control box is available with $100 upgrade

WarrantyOne-year limited warranty with lifetime support

Application 

  •      • 
    Adjust the
     SP1 and SP2 parameters of the digital pressure gauge.
  •      •  Here are the tips: Long press to increase the value; Press once, wait for 3 seconds, and then long press to
            decrease the value; Press ADJ to return to the main menu.
  •      •  Only the rubber needle included in the package is allowed to use for adjusting the set point SP1 and SP2.
            The warranty does not cover a defect resulting from misuse or negligence of this tool.