Power Supply | AC 100-120V or AC 208-240V can be selected in options bar (50/60Hz).
|
Current | 3.5-5.6A |
Refrigerant & Cooling Capacity
| • Refrigerant: R134a (~300g)- • Capacity: 722Kcal/h; 2866Btu/h; 840W
|
Temperature Control | 5~30 °C |
Temperature Accuracy | ± 0.3 °C |
Max. Pump Head | 10 meters (14psi) |
Pipe Connectors | • Φ 12mm PTC connector- • 8mm I.D, 12mm O.D. polyurethane tube is included
|
Water Flow Rate | 16 Liters / minute
|
Water Tank | • 6 Liters Water Tank made of stainless steel- • Water chiller should not be connected to a system with the capacity more than the tank's
capacity itself.
- • Use coolant instead of water for longer service life.
- • Do not use deionized water as the cooling agent, DI water may cause the copper pipe to
corrode, mainly pitting, and leak over time.
|
Water Filter | • One water filter is included for immediate use.- • Replace water filter every 3 to 6 months based on water quality and usage.
Please click on pic to order
|
Protection | Protected from overflow alarm and over temperature |
Noise Level | 85dBA (standing next to the unit) 75dBA (standing at 1 meter away from the machine)
|
Overall Dimensions | 550 X 280 X 430 mm (L x W x H) |
Shipping Weight | 100 lbs |
Operation Instructions | |
Warranty | One year limited warranty with lifetime support
|
Compliance | • CE approval- • MET Certification (UL 1450) is available upon request at extra cost, please contact our sales
representatives for a quote.
- •
|
Application Notes | • Must use Rust Protection coolants to prevent rust and corrosion. The use of tap water is prohibited.- • Please connect the chiller water outlet to the equipment water inlet, and the chiller water inlet
to the equipment outlet. Otherwise, the water flow sensor on the equipment may fail to work due to wrong water flow direction - • If you need temperature below 20C, please set Parameter F9 to the desired temperature
(10C-20C).
|