Power | - Voltage: 208 - 240VAC , 50/60Hz, single phase
- Power:9.0 KW ( 50A air breaker is required )
|
Heating furnace
| - Fibrous alumina for thermal insulation
- Double-layer steel structure with air cooling
- Vertical Split door design for easy installation of tube
- Separated temperature control unit
- Three heating zones with 1200 mm total length ( 400 + 400 + 400 mm) with independent temperature control
- Max. working temperature: 1000ºC
- Max heating rate: 20ºC/minute
- Optional: Pre-heating furnace for gas is available at the extra cost to provide heated gas to processing vessels upto 700oC va 310S alloy tube
|
Processing tube
| - Nickel base super-alloy tube with the optional diameter and CF flange
- φ30 OD x φ15 ID X 1400 L (mm )
- φ80 OD x φ40 ID X 1400 L (mm )
- Max. working pressure is dependent on max working temperature
- ≤700ºC, Pressure: ≤ 25 MPa
- ≤ 800ºC, Pressure: ≤ 20 MPa
- ≤ 900ºC, Pressure: ≤ 10 MPa
- ≤ 1000ºC, Pressure: ≤ 5 MPa
- ≤ 1100ºC, Pressure: ≤ 3 MPa
- One mechanical pressure gauge pre-installed on top of the tube (Pic. 1)
- The pre-installed Safety relief valve will turn on automatically once the pressure reaches 0.3Mpa.
- One Ni alloy mesh is installed inside the tube center to carry two layers of materials (Pic. 2 &3)
Pic. 1 Pic. 2 Pic.3 |
Temperature Controller
| - Three digital temperature controllers are installed to control 3 heating zone independently
- 30 programmable segments for precise control of heating rate, cooling rate, and dwell time.
- Built-in PID Auto-Tune function with overheating & broken thermocouple protection.
- Over-temperature protection and alarm allow for operation without attendant(s).
- +/- 1 ºC temperature accuracy.
- Eurotherm 3000 precision temperature controller is optional with +/- 0.1ºC accuracy
|
Dimensions | L 1200 x W 700 X H2500 (mm) |
Net Weight | ~150 Kg
|
Warranty | One-year limited warranty (Consumable parts such as heating elements, processing tubes, and o-rings are not covered by the warranty, please order the replacement at related products below.) |
Laptop & software (Optional)
| - Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial)
- Labview Based Temperature Control System (EQ-MTS01) enables the user to edit temperature profiles, manage heat-treatment recipes, record and plot data for MTI furnaces
|
Compliance
| - CE Certified
- NRTL or CSA certification is available at an extra cost
|