Furnace Structure
| - ZrO2 ceramic liner can stand at temperature up to 2000°C
- Kathal 1900 grade MoSi2 heating element with 1800oC max working temperature
- Double shell steel case with a cooling fan to keep outside surface cool
- 25mm dia. Saphire window is built for sample observation by optical scope or high-speed camera
- 20 mm Dia. the feeding port is built on top of the furnace for adding doping materials or an optional stirring mechanism during melting
-
|
Dimensions
| - Heating Chamber: 160 x 150 x 150mm (6.3" x 5.9" x 5.9"), 3.6L
- Overall: 515 x 500 x 730 mm (20.28" x 20" x 28.75")
|
Power | - AC 208V-240V single phase, 50/60 Hz
- 6.0 KW (50A air breaker required)
- 10 feet long 10-3 AWG heavy duty UL approved power cable is included
- Extra length cable can be ordered by click the picture below
- The plug is not included. You can order from your local hardware store according to the female plug in your lab, or order from the link below (30A plug)
- (click here to order plugs from us)
|
Working Temperature | - 1800° C (< 1 hours)
- 1750° C (continuous)
|
Heating Rate | - 0~10°C / min (suggestion =< 5°C)
|
Temperature Uniformity | - +/- 5°C over 120 mm (5") @ 1700°C
- +/- 2°C over 80 mm (3") @ 1700°C
|
Heating Elements | Kanthal Super-1900 Grade MoSi2, 4 pcs |
Thermocouple | B-Type (Pt-Rh to Pt-Rh) with Alumina tube |
Temperature controller
| - Eurotherm 3000 Series Programmable Temperature Controller.
- 1 program, 24 segments programmed with ramping, cooling, and dwelling steps
- Temperature Accuracy: ± 0.1ºC
- RS485 port is built-in for PC operation
- Operation software is included for PC operation.
- Brand new laptop with Microsoft Windows 10 with pre-installed software is available upon request.
|
Net Weight | 146 Kg |
Shipping Weight | 400 lbs |
Shipping Dimensions | 48" x 40"x 48" |
Warranty
| - One year limited warranty with lifetime support. (Consumable parts such as sample plates and heating elements are not covered by the warranty, please order the replacement at related products below.)
- ATTENTION: Any damages caused by the use of corrosive and acidic gases are not under the coverage of MTI One Year Limited Warranty.
|
Compliance | - CE Certified
- NRTL or CSA certification is available at the extra cos
|
Application Notes | - For longer service life, we recommend the heating rate not to exceed 10 °C /min.
- The cooling rate shall also not exceed 10 °C /min.
- Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with ZrO2 coating.
- Please click here to learn how to DIY a gas purging system.
- The refractory block must be inserted before closing the door.
|
Operation Instructions | |