Furnace Structure | • Two independent temperature controllers to control each zone's temperature.- • Double layer steel casing with air cooling keeps furnace surface temperature lower than 60°C.
- • High purity fibrous alumina insulation for maximum energy saving.
- • The special refractory coating on liner and heating element help increase furnace service life.
- • Split cover with Interlock protection, cut the power if door open during furnace heating process.
|
Power Requirement | - • Single Phase 208~240VAC, 3KW (20Amp circuit breaker is required)
|
Heating Zone Length | • Two zones: each zone length is 8" (200 mm)- • Total 16" (400 mm) heating zone
|
Constant Temp. Zone | • 150mm (+/-1°C) for each heating zone- • A longer constant temperature zone ( about 250mm ) can be created by setting both zones at the same
temperature - • Please click the picture in below to see detail temperature uniformity profile at 800°C.
|
Temperature Range | 100~1100°C (Continuous); 1200°C (<1 hour) |
Max. Heating Rate | ≤ 20°C /min |
Revolving Feeding System | • One thermocouple connects to the crucible from the left flange (Pic.1). The stepper motor drives the crucible to travel through the tube together with the thermocouple for real-time temperature monitoring.- • Four crucibles can be simultaneously placed into the revolving feed system and sequentially delivered to
the quartz tube under vacuum or inert gas (Pic.2,3&4).
Pic.1 Pic.2 Pic.3 Pic.4
|
Crucible Movement Control | • The crucible moving speed and distance are controlled by a stepper motor.- • Crucible travelling distance: 1mm ~ 200mm
|
Temperature Controllers | • 2 MET certified digital controllers with 30 programmable segments for precise control of heating rate, cooling rate and dwell time.- • Built in PID Auto-Tune function with overheating & broken thermocouple protection.
- • Over temperature protection and alarm allows for operation without attendant(s).
- • +/- 1ºC temperature accuracy.
- • Two K-type thermocouples (one for each zone).
- • RS485 Communications Port.
- • Optional:
- º The furnace can be operated by PC with MTS02-Y Temperature control software package
(include 2 communication chips and one 15" laptop). - º The new version of PC module is now compatible with Lab View 05/08/2014
|
Heating Element | Fe-Cr-Al Alloy doped by Mo |
Tube Accessories | • One fused quartz tube (60mm O.D. x 54mm I.D. x 1000mm L) is included.- • Two fibrous ceramic tube blocks are included for blocking heat radiation from inside tube.
(Ceramic blocks must be fully inserted into the furnace chamber before heating) - • Quartz block (Optional) is ideal for more cleanness and higher achievable vacuum, available at extra cost.
|
Digital Vacuum Gauge
|
- • Corrosion-proof Capacitance Diaphragm Gauge with ceramic coated sensor head specially designed for
vacuum measurements with the presence of corrosive gases. - • Provides a wide range (3.8E-5 ~ 1125 torr) of measurement.
|
Vacuum Pump
| |
Gas Flow Control System
| |
Water Chiller
| |
Dimensions
| 2400mm(L) x 600mm(W) x 1300mm(H)
|
Warranty | One year limited warranty with lifetime support. (Consumable parts such as processing tubes, o-rings are not covered by the warranty, please find the replacements at related products below.)
|
Laptop, software & WiFi Control (Optional) | |
Compliance | • CE Certified- • All electric components ( >24V) are UL / MET / CSA certified
- • The furnace is ready to pass TUV(UL61010) or CSA certification at extra cost.
(please click marks below to learn details)
|
Application Notes
Click here to learn the installation of a gas regulator.
| |
Operation Instructions
| |