Furnace Structure
| - Using one of the MTI split tube furnaces with 1600oC working temperature via MoSi2 heating elements.
- Heating zone length 390 mm with 135 mm constant temperature zone within +/- 1℃
- The temperature range is optional from 1200 ℃ to 1600 ℃ with +/- 0.1℃ accuracy
- Precision electronic balance with 0.001 g resolution.
- The vacuum level can reach 10-5 torr via turbopump
|
Working Temperature | - 900 ℃ - 1600 ℃ under oxygen & inert gas atmosphere ( PSIG 0.1 > 3 )
- 900oC - 1500oC under vacuum
|
Tube Size and Materials
| - High purity alumina tube with a dimension of OD60x ID54x L1200 (mm)
-
|
Force Sensor
| - High precision force sensor with 0.05% accuracy
- Resolution: 1 mg
- Sample amount: 1 g minimum; 100 g maximum. ( based on Fe powder )
- Force Sensor output data will connect to PLC or PC to show the weight change curve with temperature
|
Crucibles
| - Five alumina crucibles are included with 36 ml capacity ( OD40xID35xH40mm)
- Please click Pic. 1 to order more
|
Temperature Controllers
| - Programmable temperature controller ( Europe there for precise control of heating rate, cooling rate and dwell time.
- Built-in PID Auto-Tune function with overheating & broken thermocouple broken protection.
- Over-temperature protection and alarm allow for operation without attendant(s).
- +/- 0.1 ºC temperature accuracy.
- One K-type thermocouple.
- RS485 Communications Port.
- One brand new laptop with pre-installed software is included
|
Vacuum | - A KF40 vacuum port is built-in to connect a vacuum pump
- Achievable vacuum level depends on the selected vacuum pump
- Click the Picture below to order a vacuum pump separately
|
Gas Delivery (optional) | - 1/4" gas inlet and out are built-in with an optional mass flow controller. The recommended gas flow rate is <200 sccm.
- You may connect the TGA furnace with MTI multi-channel gas delivery system for mixing gas operation, or oxygen or a gas analyzer to the gas outlet port
|
Warranty | One-year limited warranty (Consumable parts such as heating elements, processing tubes, and o-rings are not covered by the warranty, please order the replacement at related products below.)
|
Compliance
| - CE Certified
- NRTL or CSA certification is available upon request at extra cost.(please click the marks below to learn details
|
Dimension |
Demo Video | |