Furnace Structure
| - Using one of the MTI tube furnaces and modifying it into vertical
- The temperature range is optional from 1200 ℃ to 1700 ℃
- Click the Picture below to see furnace specifications
|
Max. Temperature | - 1200 ℃ - 1700 ℃ selectable
|
Tube Size and Materials
| - 2" OD - 3.2" OD selectable. Click the picture below to see tube specifications
- For < 1200°C Max. please choose the quartz tube
- For < 1500°C, please choose Mullite tube
- For < 1700°C, please choose the Alumina tube
|
Force Sensor
| - High precision force sensor with 0.05% accuracy.
- Sample amount: 1 g minimum; 100 g maximum. ( based on Fe powder )
- Force Sensor output data will connect to PLC or PC to show the weight change curve with temperature
|
Crucibles | - One alumina crucible is included. Please click Pic. 1 to order more
- You may click Pic. 2 to choose ZrO2, BN, or Y2O3 crucible based on your application
- The customized crucible is available upon request
- Pic. 1 Pic. 2
|
Temperature Controllers
| - Programmable temperature controller ( Eurotherm-3000 for precise control of heating rate, cooling rate and dwell time.
- Built-in PID Auto-Tune function with overheating & broken thermocouple broken protection.
- Over-temperature protection and alarm allow for operation without attendant(s).
- +/- 0.1 ºC temperature accuracy.
- One B-type thermocouple, which is touched to bottom of crucible and control temperature
- RS485 Communications Port.
- PC control software and interface module are included
|
Vacuum | - A KF25 vacuum port is built-in to connect a vacuum pump
- Achievable vacuum level depends on the selected vacuum pump
- Click the Picture below to order a vacuum pump separately
|
Gas Delivery (optional) | - 1/4" gas inlet and out are built-in with an optional mass flow controller. The recommended gas flow rate is <200 sccm.
- You may connect the TGA furnace with MTI multi-channel gas delivery system for mixing gas operation, or oxygen or a gas analyzer to the gas outlet port
|
Warranty | One-year limited warranty (Consumable parts such as heating elements, processing tubes, and o-rings are not covered by the warranty, please order the replacement at related products below.)
|
Compliance
| - CE Certified
- NRTL or CSA certification is available upon request at extra cost.(please click the marks below to learn details
|
Dimension
| |
Demo Video | |
Note | - When you require a quotation, please specify
- Max working temperature?
- Max. sample weight?
- Working atmosphere?
|