400-1250ºC Max. Muffle Furnaces

1200C Muffle Furnace (12x12x12", 27 L) w/ Programmable Controller- KSL1200X-M

1200C Muffle Furnace (12x12x12", 27 L) w/ Programmable Controller- KSL1200X-M

기본 정보
Product Name 1200C Muffle Furnace (12x12x12", 27 L) w/ Programmable Controller- KSL1200X-M
Sale Price Call for Price
Product Code KSL1200X-M
Quantity 수량증가수량감소
상품 옵션
 
KSL-1200X-M-UL is a muffle furnace that offers a 12" x 12" x 12" high purity fibrous alumina chamber, gas inlet and outlet ports, and a 30 programmable segments temperature controller and UL / MET / CSA Certified electric components. It is the ideal muffle furnace for preparing samples with temperatures up to 1100 °C in research and dental labs with maximum energy saving.

SPECIFICATIONS

Structure

  • High-purity fibrous alumina insulation for maximum energy saving.
  • Heating elements on all three sides of the interior chamber for fast heating and enhanced temperature uniformity.
  • Double-layer steel structure with air cooling fan to keep furnace's exteriors safe to touch.
  • The safety interlock will automatically power off the furnace when the door is opened.
  • Built-in vent port (top) to remove undesirable contaminants and moisture and extend the lifespan of the heating elements and furnace.
Chamber Dimensions

  • Size: 12" x 12" x 12"  ( 304 x 304 x 304 mm)
  • Capacity:  27 liters (1728 cubic inches)
  • Observation window of quartz ( 1/2" - 1" diameter ) is available upon request
  •             

Power

  • 208~240V AC, Single phase, 50/60 Hz
  • 5.5 kW, 30 A. 
  • Note: A power cord is included without a power plug. 
    A proper power plug is available via the figure on the left. 
Max Operating Temperature1200 °C continuously
Temperature Stability
±1 °C
Temperature Uniformity
  • +/- 5 °C over 200×200×200 mm  @ 800°C
  • +/- 2 °C over 80×80×80 mm @ 1100 °C
Max. Heating Rate10 °C/min
Temperature Controller
  • FA-YD518P-AG temperature controller is included.
  • Proportional–integral–derivative control (PID control) and auto-tune function
  • 30 segments programmed with ramping, cooling and dwelling steps
  • Built-in over-temperature alarm and thermocouple failure alarm
  • +/- 1 ºC temperature control accuracy
  • Default DB9 PC communication port
Temperature controller (optional)
  • MTS02-Y Temperature control software kit (for YD518P series controllers) + 15" laptop package is available at the options bar. Click for detailed information about the MTS02-Y software kit. 
  • You may upgrade the temperature controller to the Eurotherm Temperature Controller which is fully compatible with LabVIEW. This package includes a Eurotherm programmer, a communication cable (RS485 - USB) and a software CD. Click the pic below to view the full spec of the Eurotherm temperature controller.       
Laptop, software & WiFi Control (Optional)
   
  • Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial) for immediate use.
  • Labview Based Temperature Control System (EQ-MTS01) enables the user to edit temperature profile, manage heat-treatment recipes, and record and plot data for MTI furnaces.
  • A wireless remote control provides up to 300 meters of operating range.
  • The above features are available upon request at an extra cost. Please contact us for more information.
Gas Fitting & Exhaust

  • One 1/4" barbed Gas Inlet was installed on the back panel of the furnace. 
  • One gas exhaust port is built on the top of the furnace casing.
     
Product Dimensions

  24" x 34" x 24" (D x H x W)   Click here to order a heavy-duty mobile cart to carry dual furnace
Shipping Dimensions48" x 40" x 50" (D x W x H)
Shipping Weight
320 lb
Warranty
  • One-year limited warranty with lifetime support.  
  • ATTENTION: Any damages caused by the use of corrosive and acidic gasses are not under the coverage of MTI One Year Limited Warranty.
Compliance
 
  • CE Certified
  • NRTL  (UL61010) or CSA certification is available upon request at extra cost. 




    Application Notes
    • For longer furnace service life, we recommend the heating rate not to exceed 10 °C /min.
    • The cooling rate shall also not exceed 10 °C /min.
    • This furnace lacks vacuum sealing, and it is not recommended to use toxic or explosive gasses in this furnace without necessary safety controls and supervision. 
    • Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with alumina coating.
    • You may order an alumina crucible by click the picture below.
             click here to choose a crucible



    Operation Instructions